Dissolved Oxygen
Range 0.00 to 90.00 mg/L (ppm) concentration; 0.0 to 600.0 % saturation
Resolution 0.01 mg/L (ppm); 0.1 % saturation
Accuracy Refer to individual specifications for probe used
HI-7641133 DO Probe Specification
Probe Body Material ABS
Smart Cap™ material Polypropylene
Cable jacket material PVC
Cable Length 1m
probe guard Stainless steel
Temperature Measurement Thermistor
Pressure 20m (29 PSI)
Probe Dimensions with guard 174 x 25mm
Response Time 45 seconds
probe ingress protection rating IP68
Sensor type Optical; luminescence quenching
HI-764833 Polarographic probe
Range 0.00 to 90.00 mg/L (ppm) concentration; 0.0 to 600.0 % saturation
Resolution 0.01 mg/L (ppm); 0.1 % saturation
Accuracy ±1.5 % of reading ±1, least significant digit
Barometric Pressure
Range 450 to 850 mmHg; 600 to 1133 mBar; 60 to 113.32 kPa;
17.72 to 33.46 inHg; 8.702 to 16.436 psi; 0.592 to 1.1184 atm
Resolution 1 mmHg; 0.1 mBar; 0.01 kPa; 0.01 inHg; 0.001 psi; 0.0001 atm
Accuracy ±3 mmHg within ±15 % from the calibration point
±3 mmHg ±1 least significant digit
Temperature
Range −20.0 to 120.0 °C
−4.0 to 248.0 °F
253.0 to 393.0 K
Resolution 0.1 °C; 0.1 °F; 0.1 K
Accuracy Refer to specification of probe used
DO Calibration
Points One or two points at 100.0% (8.26 mg/L) and 0.0% (0.00 mg/L) automatic.
Manual using a value entered by the user in % saturation, mg/L or ppm.
Temperature User Calibration 1 point, configurable
Salinity Compensation Manual; 0.00 to 45.00 PSU or ppt
0.0 to 130.0 %