HI-93754D-25 LOW RANGE COD REAGENTS, MERCURY FREE
HI-93754D-25
This method is recommended for general purpose analysis with no chloride interference.
Low Range
COD
25 Tubes
Full Description
Specification
Print Page
Accessories
Reference
Price
Quantity
Full Description
Specification
The HI-93754D-25 is a box of 25 reagent vials for COD, low range mercury-free
*This method is recommended for general purpose analysis with no chloride interference
Each box of 25 vials is supplied with a Hanna certificate of quality. The reagents are traceable to NIST SRM® 930.
Hanna offers reagents to cover three separate ranges.
- Three measurement ranges to satisfy every need – As COD levels vary depending on the application and process measuring points. Simply choose the best range for the application: low range – 0 to 150 mg/L O2, medium range – 0 to 1500 mg/L O2 and high range: 0 to 15000 mg/L O2.
- Accurate and repeatable measurements – Hanna COD reagents have been developed in accordance with Standard Methods 5220D, USEPA 410.4 and ISO 15705:2002 methods.
- Pre-dosed vials – Hanna vials contain approximately 3mL of pre-dosed reagent. The operator just needs to add a small quantity of the sample.
- Quick and accurate measurements – With pre-dosed vials, test preparation time is dramatically reduced. There is no time-consuming reagent preparation procedure or glassware cleaning.
- Safe reagents – Hanna COD reagents are safe for operators and the environment. Vials and caps have been designed to avoid accidental reagent spills. Due to the pre-dosed reagents, the amount of chemicals and handling time is minimised.
Read More
Range | 0 to 150 mg/L (ppm) | |
Sample qty. | 2.0 mL |
Dichromate EPA Method | ||
COD - LR | ||
Range (as O2) | 0 to 150 mg/L | |
Resolution | 1mg/L | |
Accuracy | ±5 mg/L ±5% of reading | |
Pack of 25 vials | HI-93754A-25 | |
Replacement reagents | HI-94754A-25 | |
COD - MR | ||
Range (as O2) | 0 to 1500 mg/L | |
Resolution | 1mg/L | |
Accuracy | ±15 mg/L ±4% of reading | |
Pack of 25 vials | HI-93754B-25 | |
Replacement reagents | HI-94754B-25 | |
Dichromate Method | ||
COD - HR | ||
Range (as O2) | 0 to 15000 mg/L | |
Resolution | 10mg/L | |
Accuracy | ±150 mg/L ±3% of reading | |
Pack of 25 vials | HI-93754C-25 | |
Replacement reagents | HI-94754C-25 | |
Dichromate Mercury-Free Method | ||
COD - LR | ||
Range (as O2) | 0 to 150 mg/L | |
Resolution | 1mg/L | |
Accuracy | ±5 mg/L ±5% of reading | |
Pack of 25 vials | HI-93754D-25 | |
Replacement reagents | HI-94754D-25 | |
COD - MR | ||
Range (as O2) | 0 to 1500 mg/L | |
Resolution | 1mg/L | |
Accuracy | ±15 mg/L ±4% of reading | |
Pack of 25 vials | HI-93754E-25 | |
Replacement reagents | HI-94754E-25 | |
Dichromate ISO Method | ||
COD - LR | ||
Range (as O2) | 0 to 150 mg/L | |
Resolution | 1mg/L | |
Accuracy | ±5 mg/L ±5% of reading | |
Pack of 25 vials | HI-93754F-25 | |
Replacement reagents | HI-94754F-25 | |
COD - MR | ||
Range (as O2) | 0 to 1500 mg/L | |
Resolution | 1mg/L | |
Accuracy | ±15 mg/L ±4% of reading | |
Pack of 25 vials | HI-93754G-25 | |
Replacement reagents | HI-94754G-25 |