Ideal for on the spot Pool, Spa and Hot Tub water testing

.

Fosfat

836 kr Exl. moms
1 045 kr (inkl. moms)
Tillbehör
Art.nr.
Pris
Kvantitet
Beskrivning Specifikation Manuals

HI-7134 Pool Line Phosphate low range Checker HC® colorimeter: Range 0.00 to 2.50 ppm (mg/L)

With its high accuracy and low level detection at the touch of a button - the HI-7134 Pool Line Phosphate Checker is ideal for the daily care of your swimming pool, hot tub or spa.

Orthophosphates are found in natural waters and wastewaters. They are commonly added to drinking water as a corrosion inhibitor. The instantaneous analysis of orthophosphates by colorimetric determination provides rapid results using a standard analysis technique.

It is important to monitor phosphate to keep algae growth at bay and protect human health. Phosphates are also ideal nutrients for the algae that causes green water.

With the HI-7134, you can measure phosphate in a simple and affordable way. Developed as a more accurate alternative to chemical test kits, this checker provides fast results.

Main Benefits of the Checker

  • Testing is quick and easy
  • High precision results - more accurate than chemical test kits
  • Auto shut-off feature
  • Fully portable, fitting easily into both palm and pocket

Supplied with

  • 2 sample cuvettes and caps
  • Starter pack of reagents x 6 (re-ordering information HI-7134-25 x 25 tests)
  • 1 x 1.5 AAA battery
  • Instructions

Also available to buy: HI-7134-11 Calibration set

Use Reagents HI-7134-25 with the HI-7134 Pool Line Phosphate Low Range Handheld Colorimeter - Checker®HC, x 25 tests

Use Calibration Check Set HI-7134-11 with the HI-7134 Pool Line Phosphate Low Range Handheld Colorimeter - Checker®HC.

Läs mer

Range 0.00 to 2.50 ppm (mg/L)
Resolution 0.01 ppm (mg/L)
Precision ±0.04 ppm (mg/L) ±4% of reading @ 25°C
Light Source LED @ 525 nm
Light Detector Silicon photocell
Environment 0 to 50°C (32 to 122°F); RH max 95% non-condensing
Battery Type (1) 1.5V AAA
Auto-off After 10 minutes of non-use and after read
Dimensions 86 x 61 x 37.5 mm (3.2 x 2.4 x 1.5”)
Weight 64 g (2.25 oz.)
Method Adaptation of Standard Methods for the Examination of water and wastewater. 20th edition, Ascorbic Acid method. The reaction between phosphate and the reagent causes a blue tint in the sample.
PLEASE NOTE: In some cases, samples may contain other parameters which may interfere with results
Interference may be caused by: Iron above 50 mg/L
Silica above 50 mg/L
Silicate above 10 mg/L
Copper above 10 mg/L
Hydrogen sulphide, arsenate, turbid sample and highly buffered samples also interfere.