Hanna HI-4004 Calcium Half-Cell ISE

HI-4004

The HI-4004 is a liquid membrane, half-cell ion selective electrode (ISE) for the determination of calcium (Ca2+) in solution.

Liquid Membrane sensor
Kalcium
6 months warranty

11 189.26 kr Exl. moms
13 986.57 kr (inkl. moms)
Tillbehör
Art.nr.
Pris
Kvantitet
Beskrivning Specifikation

The PVC membrane comprised with organic ion exchanger produces a potential change due to the calcium ion exchange between the organic polymer membrane and the sample. The internal sensing elements are housed within a durable epoxy body.

The HI-4004 is ideal for a variety of applications in the food and beverage industry and water analysis.

  • Liquid membrane sensor
  • Detection from 0.12 to 40,080 mg/L Ca2+
  • For use with HI5315 reference half-cell
  • 1m cable and BNC connector

The HI-4004 is a liquid membrane, half-cell Ion Selective Electrode (ISE) for the detection of calcium (Ca2+) in solution. The PVC membrane comprising an organic ion exchanger produces a potential change due to the calcium ion exchange between the organic polymer membrane and the sample. The internal sensing elements are housed within a durable epoxy body.

The HI-4004 calcium ion selective electrode is a potentiometric half-cell sensor housed within an epoxy body. The internal elements are isolated from the sample by an organic polymer sensing PVC membrane. This liquid membrane sensor requires a separate reference electrode to complete the measurement system.

Hanna’s liquid membrane half-cell ISEs are designed for use with the HI-5315 reference half-cell. The reference half-cell features a cone style junction. The skirted cone geometry of the sensor forms the liquid junction with the test solution at the tip of the junction cone, producing a highly stable reference potential and consistent flow rates of electrolyte into the solution. A simple press of the top portion of the ISE makes it easy to empty the electrolyte fill solution and replace when necessary.

For the HI-4004 to measure calcium accurately, it is important that the ionic strength of the standards and sample be adjusted to a high and constant value with the calcium ionic strength adjuster (ISA) solution. The ISA ensures that the activity coefficient is constant so the concentration of free calcium ions can be measured. Measuring all solutions with a constant ionic strength reduces the margin of error between measurements. The HI-4004 calcium liquid membrane ISE can measure from 3µM (0.12 mg/L) to 1M (40,080 mg/L).

Läs mer

ParameterCalcium
TypePolymer Membrane; Half-cell
Measurement Range1M to 3X 10-6M 40080 to 0.12 ppm
Optimum pH Range4 to 10
Temperature Range0 to 40°C
Approximate Slope+28
Body O.D.12 mm
Insertion Length120 mm
Body MaterialPVC
Cable1m coaxial
ConnectorBNC
Possible ApplicationsDetermination of free calcium in beverages, water, and seawater