HI-4104 CALCIUM COMBINATION ISE ELECTRODE
HI-4104
The HI-4104 is a calcium combination ISE ideal for the detection of free Calcium in food, beverages, water, and seawater.
The HI-4104 is a liquid membrane, combination ion selective electrode (ISE) for the determination of calcium (Ca2+) in solution. The PVC membrane along with an organic ion exchanger produces a potential change due to the calcium ion exchange between the organic polymer membrane and the sample. The internal sensing elements are housed within a durable polyetherimide (PEI) body.
The HI-4104 Calcium Ion Selective Electrode is a complete potentiometric cell that contains a Silver/Silver Chloride (Ag/AgCl) reference housed within a PEI plastic body with a Potassium Chloride electrolyte. The internal elements are isolated from the sample by an organic polymer sensing membrane. Hanna’s liquid membrane combination ISEs are designed with a cone style junction.
Before testing, it is important that the ionic strength of the standards and sample be adjusted to a high and constant value with the ionic strength adjuster (ISA) solution (HI-4000-00). The HI-4104 calcium liquid membrane ISE can measure from 3µM (0.12 mg/L) to 1M (40,080 mg/L).
BNC Connection – BNC is one of the most commonly used connector types in pH meters and electrodes. Compatible with any meter or electrode that uses BNC connectors.
Läs mer
| Parameter | Calcium | |
| Type | Polymer Membrane; Combination | |
| Measurement Range | 1M to 3X 10-6M 40080 to 0.12 ppm | |
| Optimum pH Range | 4 to 10 | |
| Temperature Range | 0 to 40°C | |
| Approximate Slope | +28 | |
| Body O.D. | 12 mm | |
| Insertion Length | 120 mm | |
| Body Material | PEI/PVC | |
| Cable | 1m coaxial | |
| Connector | BNC | |
| Possible Applications | Determination of free |